Hanyoung GR200 paperless chart recorder
Hanyoung GR200 paperless chart recorder
Hanyoung GR200 paperless chart recorder
Hanyoung GR200 paperless chart recorder

Hanyoung GR200 paperless chart recorder

Hanyoung GR200 paperless chart recorder

Description

· Suffix code

Model Code Information
GR200 - Graphic recorder 145(W) x 145(H) x 173.5(D) ㎜
Input channel 2     2 channel
4     4 channel
8     8 channel
12     12 channel
External contact
input & output
(DI / DO)
N   None
1   DI 2 contact + DO 6 contacts
2   DI 4 contact + DO 12 contacts
Communication function 0 RS422/485
1 Ethernet

· Specification

[ Range configuration for the input types ]
Input type Measurement range(°C) Measurement range(°F) Degree
Thermoresistor (RTD) Pt100Ω Pt-0 -200 ~ 640 -300 ~ 1180 ±0.15 % of F.S ±1digit
Pt-1 -1 -200.0 ~ 640.0 -300.0 ~ 1180.0
Pt-2 -200.00 ~ 640.00 -300.0 ~ 1180.0
KPt100Ω KPt-0 -200 ~ 500 -300 ~ 1000
KPt-1 -200.0 ~ 500.0 -300.0 ~ 1000.0
KPt-2 -100.00 ~ 150.00 -300.0 ~ 1000.0
Thermocouple (TC) K K-0 -200 ~ 1370 -300 ~ 2500 ±0.15 % of F.S ±1digit
K-1 -200.0 ~ 1370.0 -300 ~ 2500 ±0.15 % of F.S ±1digit(*2)
J -200.0 ~ 1200.0 -300 ~ 2300 ±0.15 % of F.S ±1digit(*2)
E -200.0 ~ 1000.0 -300 ~ 1800 ±0.15 % of F.S ±1digit(*2)
T -200.0 ~ 400.0 -300 ~ 750 ±0.15 % of F.S ±1digit(*3)
R 0.0 ~ 1700.0 0 ~ 3100 ±0.15 % of F.S ±1digit(*4)
B 0.0 ~ 1800.0 0 ~ 3300 ±0.15 % of F.S ±1digit (*1)
S 0.0 - 1700.0 0 ~ 3100 ±0.15 % of F.S ±1digit(*4)
L -200.0 ~ 900.0 -300 ~ 1300 ±0.15 % of F.S ±1digit(*2)
N -200.0 ~ 1300.0 -300 ~ 2400 ±0.15 % of F.S ±1digit(*3)
U -200.0 ~ 400.0 -300 ~ 750 ±0.15 % of F.S ±1digit(*3)
Wre 5-26 0.0 ~ 2300.0 0 ~ 4200 ±0.15 % of F.S ±1digit(*4)
PL-Ⅱ 0.0 ~ 1390.0 0 ~ 2500 ±0.15 % of F.S ±1digit
VDC -10 - 20 ㎷ -10.00 ~ 20.00 ±0.15 % of F.S ±1digit
* The current input (4~20㎃ DC)
is available when you use resistance 250Ω (0.5W/0.1%)
on input terminals
0 - 100 ㎷ 0.00 ~ 100.00
1 - 5 V 1.00 ~ 5.00
0 - 30 V 0.00 ~ 30.00
 (*1) 0 ~ 400℃ : ±5% of F.S ± 1digit  (*3) -200 ~ -150℃ : ±0.4% of F.S ± 1digit
 (*2) -200 ~ -150℃ : ±0.2% of F.S ± 1digit        -150 ~ -100℃ : ±0.2% of F.S ± 1digit
   (*4) 0 ~ 200℃ : ±0.2% of F.S ± 1digit

[ Specification for the input ]
VDC Thermocouple (TC) K, J, E, T, R, S, B, N IEC 584
PL-Ⅱ, Wre 5-26 ASTM E988
U, L DIN 43710, IEC 751
Thermoresistor (RTD) Pt100 Ω DIN IEC 751
KPt100 Ω JIS C1604-1989 (OLD)

[Power input]
Power voltage 100 - 240 V AC Voltage fluctuation rate ±10 %
Power frequency 50 - 60 ㎐
Power consumption 22 VA max
Maximum internal fuse ratings 250 V AC
Internal voltage Primary terminal* and secondary terminal** : Minimum 1500 VAC for 1 minute
Primary terminal* and FG terminal : Minimum 1500 VAC for 1 minute
Secondary terminal** and FG terminal : Minimum 1500 VAC for 1 minute
Secondary terminal** and secondary terminal** : Minimum 500 VAC for 1 minute
* Primary terminal : Power terminal (except the FG terminal) and the relay output terminal
** Secondary terminal : Sensor input terminal, contact input terminal, communication terminal
Insulation resistor 20 ㏁ between the power terminal and the FG terminal or 500 VDC

[Sensor input]
Number of channels 2, 4, 8, 12 (Refer to the type configuration)
Input type 2 thermoresistors (Pt-100, KPt-100) 12 thermocouples (K, J, E, T, R, B, S, L, N, U, Wre 5-26, PL-II) 4 VDC (-10 - 20 mV, 0 - 100 mV, 1 - 5V, 0 - 30 V)
Sampling period 250 ㎳
Current to measure the thermoresistor (RTD) About 0.21 ㎃
Input resistor Thermocouple : More than 1 ㏁, VDC : More than 1 ㏁
Allowable wiring resistor Thermoresistor : Maximum 100Ω/wire (The RTD is up to 10Ω/wire for the range of -100.00 - .150.00) Thermocouple : Less than 200 Ω VDC : Less than 2 ㏀
Impact of the wiring resistor Thermoresistor : ±0.3 °C/10 Ω (The 3 lines have the same wiring resistors)
Allowable input voltage Thermocouple : Less than ±10 V DC, VDC : Less than ±33 V DC
Degree ±0.15 % of F.S, ±1 digit (Except the RJC temperature error) ※ Refer to the input table
The error in the reference junction compensation (RJC) ±2.0 ℃ (0 ~ 50 ℃)
Sensor short detection (Burn-out) UP-Scale for the short

[Contact input]
Maximum input 4
Input method No voltage contact input
On/Off detection resistor Consider on less than 1 ㏀ minimum and off larger than 10 ㏀
Minimum detection time 0.25 second

[Alarm output]
Maximum number of outputs 12
Output type Relay output
Maximum ratings 5 A 250 V AC, 5 A 30 V DC
Recommended ratings 3 A 250 V AC, 3 A 30 V DC
Relay life 50,000 times at the maximum ratings, 100,000 times at the recommended ratings

[Display specification]
Display TFT color LCD
(115.2 × 86.4 ㎜, resolution : 640 × 480 pixel, LED Backlight)
Backlight life 40,000 hours

Language in use

Korean, English, Chinese (simplified)

[Memory specification]
Save function Internal memory - volatile memory : 900 KB - Save up to 4 hours with the interval of 1 second - Non-volatile memory : 80 MB . Save up to 15 days with the interval of 1 second External memory - SD card (8GB) : Save up to a 4 years with the interval of 1 second ※ Support the SDHC
Save period User configuration (1, 2, 5, 10, 20, 30, 60, 120 seconds)
Memory information Measurement from each channel, Burn-out, DI (contact input), ALARM, relay output status

[Use environment]
Ambient temperature 0 ~ 50 °C
Temperature change 10 °C/h or below
Ambient humidity 20 ~ 90 % RH (no condensation)

[Storage environment]
Ambient temperature -20 ~ 70 °C
Temperature change Less than 20 °C/h
Ambient humidity 5 - 95 % RH (no condensation)

[Impact from the ambient temperature]
VDC, thermocouple sensor ±0.01 % of F.S / °C
Thermoresistor sensor ±0.06 °C/°C

Product Details

Contact us

optional