Spectra A1/L1 - Specifications
Performance
Accuracy (complete system at 25ºC) +/- 1ºC(+/- 2ºF)
Sensitivity 0.1ºC (0.2ºF)
Repeatability +/- 0.1ºC (+/- 0.2ºF)
Hysteresis None
Measurement range Dew/frost point meas. of -60º to +50ºC(-76ºF to 120ºF)
Required sample flow 0.5 - 5.0 scfh (.25 - 25 L/min)
Maximum cooling rate 0.5ºC/sec (1ºF/sec)
Display resolution 0.1ºC/°F (dew point)
 
Functional
Output
Analog 4-20mADC, 750 ohm max., isolated
Standard factory scaling -60º to +50ºC (-76º to +120ºF). Field scaleable via front panel switch or serial port
Resolution 0.025ºC (0.5ºF) at standard scaling
Alarm 1 Form C, 10 amps, 240 VAC/24/VDC, resistive load. Adjustable dead-band. Enable/Disable via front panel switch or serial port. Field-scaleable via front panel switch or serial port.
Digital
Serial RS-232C (DB-9 connector); RS-485 (Two wire), isolated
Status Lines Two 3 mA logic-level outputs, isolated Standard (factory) assignments: Control (data valid), Dirty Mirror Field-selectable for Control (data valid), Dirty Mirror or Alarm condition
Loop Power Supply 28 VDC, nominal, 40 mA max., isolated (for additional, user-supplied measurements)
Display 2 lines x 16 characters, LCD
Indicator Power-on LED
Inputs (for additional, user-supplied optional measurements)
Analog 4-20mADC, isolated, 7.5 volt compliance required. Field-scaleable via front panel switch or serial port
Power
Standard 115 VAC (+10%/-15%), 230 VAC (+15%/-20%), 47-63 Hz, 85 watts maximum
Optional 100 VAC (+20%/-15%), 200 VAC (+15%/-20%), 47-63 Hz.
 
Operating Ranges
Ambient/Coolant temp. (Model L1 only) 0º to 50ºC (32º to 120ºF)
Process pressure -7.5 to 175 psig (0.5 to 11.5 bar ABS)
Coolant pressure (Model L1 only) 30 psig (2 bar) max.
 
Physical
Weight 5.9 kg. (13 lb.)
Shipping Weight 6.7 kg. (15 lb.)
Sample Connections, Sensor 1/4" compression (1/16" NPTF)
Auxiliary cooling (mandatory on Model L1 only)
Standard water cooling jacket
Connections 3/8" hose adapter (1/8" NPTF)
Cooling Water 2 L/min (0.5 gal/min)
Environmental
Enclosure Designed to exceed NEMA-12, IP 65 requirements
PT-300A Pressure transducer
 
Accessories
Tool Kit (standard) Soft, fabric pouch containing screwdriver, wrench, fuses, O-rings, cleaning solution/applicators, and instruction card.
Options
SSM 1 Sample system module containing vacuum pump and flow control valve in plastic housing
SM1 Sample module with pump and oil bath filter; atm pressure only
SM2 Sample module with pump and sintered steel filter; atm pressure only
SM3 Sintered steel filter; pressure to 125 psig
OBF Oil bath filter; atm pressure only
BF10DX Particulate and liquid oil filter, clear bowl; pressure to 150 psig
FM1 Flow meter with control valve; 0-5 scfh, 100 psig max.