MMY 2650 - Specifications
Trace Moisture Analyzer

Probe
Sensor Planar gold/aluminum capacitance
Calibration Range -80°C to +20°C (-112°F to +68°F)
special calibration to -100°C (-148°F)
Recommended Recalibration Cycle 6-12 months depending on the application
Accuracy

+/- 2°C ( +/- 3.6°F) in the standard calibration range

Repeatability Better than 1°C ( +/- 1.8°F)
Calibration Data Stored in analyzer’s EPROM microprocessor
Standard Operating Temperature -20°C to +60°C (-4°F to +140°F)
Max. Relative Humidity 50% @ dew points > 0°C (32°F)
Temperature Sensor Zener device, range -70°C to +70°C (-94°F to +158°F)
Signal Transmission Frequency, generated by probe electronics
Operating Pressure 0 to 1750 psig (0 to 120 bar)
Recommended Flow Rate 1 to 5 SCFH (in a bypass mode)
Gas Flow Velocity Static up to 165 ft/sec @ 14.7 psi, 16.5 ft/sec @ 150 psig,
1.65 ft/sec at 1500 psig with no particles in the stream; higher with sintered end cap
Probe Tube 1/2” diameter, 316 stainless steel
Mounting Adapter 1/2” tube x 1/2” MNPT, 316 SS compression fitting standard
Cable Connection Rugged multi-pin connector; screw terminal in explosion-
proof junction box with 1/2” FNPT conduit connection
Probe Cable 4-conductor, AWG 22, stranded, shielded to maintain EMI/RFI/ESD resistance, up 1000 ft., 5 ft. standard
Weight .45 kg. (1 lb.)
 
MMY 2650 Analyzer
Electronics State of the art micro-controller providing utmost flexibility to meet application needs
Standard Inputs 2 (moisture and temperature)
Optional Input For pressure transducer providing live pressure compensation or other sensor. Signals: 0/1 to 5V, 0/4 to 20mA loop powered, or 4 to 20 mA. If live measurement is not available, pressure compensation can be achieved by entering a constant pressure value in matrix location V3HO
Moisture Probe Interconnects with DY 5 probe
User Interface Five push-buttons, easy configuration using a matrix
Display Alpha-Numeric LCD, displays measured value with units of measure, matrix location and programming instructions, error indication with error code if malfunction occurs; user selectable scanning feature alternating the display every 5 seconds through active channels (3 max)
EMI/RDI/ESD Protection Full compliance with EN 61326-1
Units of Measure (Moisture) Dew point °C, °F, ppmv (nneds pressure emasurement using the optional input for live calculation or pressure constant entered in matrix location V3-HO), lbs/MMSCF, g/m3, g/kg, vapor pressure in hPa, mmHg, rh%, process pressure calculated dew point °C, °F, (needs temperature measurement using the optional input for live calculation or pressure constant entered in matrix cell V3-HO))
Units of Measure (Temperature) °C, °F
Units of Measure (Pressure) Optional input used with a pressure transducer: bara, barg, psia, psig, hPaa, hPag
Meaurement Ranges User Programmable
Analog Outputs 3, each configurable to any input, 0/4 to 20 mA, load resistance <500 Ohms, 0/1 to 5V, source resistance 249 Ohms, user selectable range, user selectable condition in case of error to 110%, -10% or hold at last measured value
Digital Outputs 4 relays (SPDT dry contacts rated at 250V AC, 2.5 A, PAC = 300VA, cos phi > 0.7, P DC 100W, 100 VDC). 1 relay is system alarm. 3 relays are configurable to any input failsafe mode: energized/de-energized selectable, programmable hysteresis, high/low alarm selectable
Serial Output RS 485, update rate once per second
Serial Communication RS 485, needs GEI communication software for setup or diagnostics
Program Non-volatile memory
Data EEPROM
Oper./Storage Temp -10°C to 50°C (14°F to 122°F)
Supply Voltage 85 to 275 VAC, optional 18 to 36 VDC
Power Consumption 5.8 VA for line voltage units, 2.2 W for DC powered units
Enclosures Wall mount, IP54, NEMA 12, separate connection compartment
Cable Entry PG cable glads PG 9, 2 x PG 11, PG 13
Weight 1 kg (2.2 lbs)
 
Panel Mount Enclosure Specifications
Material Black anodized aluminum
Dimensions 144mm x 144mm panel (5.67” x 5.67”)
Depth Maximum protrusion at the rear of the panel: 209mm (8.23”)
Maximum protrusion at the front of the panel: 8.25mm (0.32”) w/ bezel
Maximum protrusion at the front of the panel with door: 32mm (1.26”)
Panel cutout 138 mm x 138mm (5.43” x 5.43”)
Wiring Same configuration as the wall mount unit, wired in the rear
Mounting Insert from front into the panel, install the clamps, tighten the clamps’ screws from the rear against the panel using a long screwdriver
Front panel Overlay w/membrane buttons integrated LEDs & clear window for display