DewPro MMY 245 - Specifications
Operating Voltage Battery Powered – 4x “D” cells
Weight <7 lbs., including 4 common “D” size batteries
Display

Large, 4 digit LCD display with rate of change bar graph indicator.

Low battery indicator “Hot Start” power up feature displays most recent setup
Enclosure NEMA-4X/IP66 impact resistant plastic enclosure; intrinsically
Safe: Designed to meet FM Class 1, Division 1, Groups A-G, approval pending
11” (L) x 7” (W) x 7-1/8” (D)
(28 cm x 17.7 cm x 18 cm)
Sensor Planar
Measurement Range -100°C to +20°C (-148°F to +68°F) dew point temperature
Accuracy +/- 2°C ( 3.6°F) dew point temperature
Readout Units of Measure Dew Point °C, Dew Point °F, ppmv, Lbs/MMSCF, g/m3, g/kg, %RH
Secondary Readout Units of Measure Temperature °C, Temperature °F, Pressure psig**, Pressure psia**, Pressure bars**, arbitrary scaled input voltage, plus all primary units of measure
** With optional pressure transducer
Repeatability +/- 1.0°C dew point temperature
Max. Pressure Rating 150 psig (10 bar)
Operating & Storage Temperature Stainless Steel 1.4571 with G 1 / 2 thread (DIN/ISO 228) and O-ring seal out of Viton
Resolution 0.1°C dew point temperature
Calibration Factory recalibration recommended every 6-12 months depending on use
Wrench Width 30 mm
Protection Cap w/10um
Weight 830 grams (1.8 lbs.)