Hygrotec MMY 170 / DY 75 - Specifications
Trace Moisture Analyzer

DY 75 Probe Specifications
Sensor Planar gold/aluminum capacitance
Calibration Range -90°C to +20°C (-130°F to +68°F) dew point temperature
Recommended Recalibration Cycle 6-24 months, depending on application and required accuracy
Accuracy

+/- 2°C (+/- 3.6°F)

Repeatability +/- 1°C (+/- 1.8°F)
Max. Relative Humidity 50% @ dew points > 0°C (32°F)
Temperature Coefficient <0.2°C /°C (<.36°F/°F)
Ambient & Storage Temperature -60°C to +50°C (-76°F to +122°F)
Nominal Operating Temperature -20°C to +40°C (-4°F to +105°F)
Max. Operating Pressure 350 bar (5100 psi)
Helium Leak Rate <10-7, mbar l/s
Max. Gas Velocity 50.0 m/s at 1 bar; 5.0 m/s at 10 bar; 0.5 m/s at 100 bar
Signal Transmission Standardized pulse (PFM) via three-wire cable
Probe Mounting Adapter Stainless Steel 1.4571 with G 1 / 2 thread (DIN/ISO 228) and O-ring seal out of Viton
Rating IP 65
Tightening Torque 50 Nm
Wrench Width 30 mm
Protection Cap w/10um
Weight 830 grams (1.8 lbs.)
 
MMY 170 Analyzer Specifications
Electronics Microprocessor-controlled
Measurement Units Dew point temperature in (C, ppmv, ppmw (optional)
Instrument Range -100°C to +20°C (-148°F to +68°F), standard. 0.01 to 9999 ppmv, and ppmw, (optional)
Display 4-character LCD w/bar graph 0-100% and two control fields
Outputs Two current outputs of 0/4-20 mA into 500 ohms max.,
Two 0/2-10 V output, 10 kohm; programming range, damping, and fail-safe mode (i.e. min/max/hold); two alarm relays SPDT dry contacts rating 2.5A, max. 250 V AC, max. 300VA at cos<0.7, max. 100 W @ 100 VDC, programmable set points, and hysteresis; one system alarm, rating same as above, triggered by the self-diagnostics program. Rackbus: Serial bus connections to ZA 672 communications interface. User Interface: Six push-buttons for display manipulation, setting parameters and functions utilizing the supplied programming matrix.
Supply Power Module 24 VDC, tolerance 20 to 30 VDC, max. 2 V ripple within the tolerance, max. 125 mA
Power Supply Monorack II, 230/115 VAC, wall mount enclosure
Analyzer Module Formatted to Europa Card 100 x 160 mm, fitting into a Racksyst, electrical connection via 28 pin plug DIN 41612
Front Panel 7 pitch (35 mm), black plastic with blue overlay
Enclosures Monorack wall mount, 65 x 165 x 200 mm with integral
power supply (selectable voltage 220/110 V AC, 200/100 V AC, or 48/24 V AC); panel mount version 1, 144 x 48 x 260 mm (requires 24 V DC power supply or optional 220/115 V AC power supply; this version supports up to four Hygrotec MMY 170 analyzers with one NT 471 power supply; cable connection via screw terminals.
Protection IP 20
Weight Electronics module: 300 grams (.65 lb.)
Monorack: 1.3 kg (2.86 lbs.)