| Sensing
Element |
Resistance
change of bulk polymer |
|
Accuracy at 25º (77º F): |
+/-
2%, 3% or 5% RH at 20 to 95% RH, includes hysteresis,
linearity and repeatability |
| Operating
Range |
0%
to 99% RH (noncondensing), -29º to +60º
(-20º to +140º) |
| Long
Term Stability |
Less
than 1% drift per year typical |
| Temperature
effect |
Less
than 0.11% per º(0.06% per º) |
| Sensitivity |
0.1%
RH |
| Repeatability |
0.5%
RH |
| Linearity |
See
accuracy |
| Hysteresis |
Less
than 1% RH |
| Signal
Outputs |
4
to 20mA, 0 to 5 volts, 0 to 10 volts, (0 to 100% RH
linear) |
| Supply
voltage |
18
to 36 VDC |
| Maximum
load (Current Output Only) |
ohm
= Supply –10VDC/.02 Amps |
| Storage
temperature |
-65º
to +70ºC (-85º to+158ºF) |
| |
| |
| Sensing
Element |
Resistance
change of bulk polymer |
| Single
point calibration range |
20%
to90% RH |
| Single
point calibration accuracy |
+/-
2% RH |
| Reference
probe accuracy |
+/-
2% RH(20% to 90% RH) |
| Operating
Range (RH Sensor) |
0%
to 99% RH (non-condensing)
-40°F to +170°F (-40°C to +76°C) |
| Operating
Range (Electronics) |
0% to 95% RH (non-condensing)
-20°F to +140°F (-29°C to +60°C) |
| Long
term stability |
Less
than 1% drift per year typical |
| Temperature
effect |
Less
than 0.11% per ºC (0.06% per ºF) |
| Probe
dimensions |
Diameter:
0.5 inches
Length: 3 inches |
| Power
supply |
9VDC
alkaline battery, continuous use for 40 hrs., Auto
power-off |
| Storage
temperature |
-85°
to +158°F
(-65° to +70°C) |
| |
| |
| Sensing
Element |
10K
thermistor at 25ºC (77ºF) |
| Accuracy
at 25ºC (77ºF) |
+/-
1.2º C(+/- 2.0ºF) |
| Signal
Output |
Direct
Connection |
| |
| |
| Sensing
Element |
1000 ohm thin film platinum RTD (.00385 alpha) |
| Accuracy
at 25ºC (77ºF) |
+/-
0.3ºC (+/- 0.5ºF) |
| Signal
Output |
Direct
Connection |
| |
| |
| Sensing
Element |
1000 ohm thin film platinum RTD(.00385 alpha) with
signal conditioning |
| Accuracy
at 25ºC (77ºF) |
+/-
0.3ºC (+/- 0.5ºF) |
| Long
term stability |
Less
than 0.2°F per year |
| Temperature
effect |
Less
than 0.02% per ºC (0.01% per ºF) |
| Sensitivity |
0.1% |
| Repeatability |
Better
than 0.1% |
| Linearity |
Less
than 0.01% |
| Sensor
Interchangability |
+/-
0.3º C(+/- 0.5 ºF) |
| Signal
Outputs |
4
to 20 mA, 0 to 5 volts, 0 to 10 volts, (-20º
to 140º linear) |
| Supply
voltage |
12
to 36 VDC |
| Storage
Temperature |
-65º
to 70º C(-85º to 158ºF) |