DewPro MDR3 - Specifications
Relative Humidity Probe

Sensing Element Silicon-based polymer, capacitance principle, IC electronics
RH Range 0 to 100%
RH Accuracy +/- 2% in the range of 0% to 90%
+/- 3% in the range of 90% to 100%
Dew Point Range -15°C to +75°C (5°F to 167°F)
Dew Point Accuracy

Better than +/- 1°C (± 1.8°F) if T>30°C (86°F) and RH>40%,T<30°C (86°F) and RH>30%

Repeatability Better than 0.5% RH
Hysteresis Less than +/- 0.9%
Standard Operating Temperature -10°C to +85°C (14°F to 185°F)
Temperature Accuracy +/- 0.5°C (+/- 0.9°F)
Maximum Operating Pressure 250 psig (17 bar)
Sensor Electronics Integrated circuitry with a platinum RTD temperature sensor
Signal Transmission Moisture and temperature frequencies allowing 1000 ft. (350 meters) of standard four-wire cable (shielded to maintain EMI/RFI/ESD resistance)
Probe Tube 316 stainless steel; 0.5” diameter; 6.8” long (12.7 mm x 172 mm)
Typical Probe Mounting 0.5” tube x 0.5” MNPT compression fitting or ANSI flange
Sensor Guard Rugged, removable, easy-to-clean, 100 micron sintered 316 L stainless steel filter; additional hydrophobic filter on the sensor element allowing the penetration of water vapor but not water droplets
Weight 0.5 lbs. (0.23 kg)
Approval CENELEC, FM (Eex ia) IIC T4, 95°C (203°F) using the Zenar barrier kit IS 20 ST from General Eastern or equivalent