Model Dew-Stat MA-99 - Specifications
Performance
Accuracy +/-1.0ºF dew point from 30º to 50ºF,
+/-2.0º dew point over full measurement range
Repeatability +/- 0.2ºC
Measurement Range +20º to +95ºF dew point (at 95ºF ambient temp.,55 psig),
+10º to +77ºF dew point (at 77ºF ambient temp., 55 psig)
Sample Flow 1 scfh (settable, 0.5 - 2.0 scfh with integral throttling valve and flow meter)
Autoclean Function Programmable interval, requires 20 psig minimum
AFAC (Anti-False Alarm Circuit) ON or OFF, DIP switch selectable under access panel.
 
Functional
Output
Alarm 1 Form C, 2 amps, 240 VAC/24 VDC resistive load, non-latching, NO/NC/COM field-settable via switch under access panel, automatic reset.
Trip Conditions measured dew point above alarm set point; sensor head left open (time delay)
Audible Alarm 85 dBA at 3 feet, variable tone, silence button on front panel, automatic reset
Indicators 3 1/2 digit LED dew point readout with º/º switchable display. LED indicators for the following conditions: Auto-clean, Out of Range, Service; Alarm Adjustable visual flow meter with graduations 0-2 scfh
Diagnostics Water phase change calibration check
Analog (opt.) 4-20mA DC, 500 ohm maximum load
Inputs
Power 115 VAC (+10% / -15%), 60 Hz, 35 watts max.
 
Operating Ranges
Ambient Temperature +32º to +120ºF
Sample Pressure 0-150 psig
Factory Setting (field resetable):
Temperature Units ºF
Alarm Point 39ºF
Autoclean Every 3 hours
AFAC Enabled
Sensor Open Alarm 20 minute delay before alarm.
 
Physical
Dimensions 12"H x 10"W x 8"D
Case Material Molded thermoplastic
Weight 10 lbs.
Connections
Air 1/8" NPTF, top of enclosure
Electrical Terminal strip, 12 gauge wire maximum, bottom of enclosure
Sample Air Vent Open, unobstructable, bottom exit
Environmental Constructed to NEMA-12 requirements
Sensor Mirror Replaceable
Mounting Vertical, Wall mount
Electrical Env. Fast transient burst per IEC 801-4; ESD per IEC 801-2; Line sure per IEC 801-5
Filter In-line, replaceable