Model M3 - Specifications
Performance
Accuracy (complete system at 25ºC)  
Dew/Frost Point +/- 0.2ºC (+/- 0.36ºF)
Temperature +/- 0.15ºC (+/- 0.27ºF)
Relative Humidity +/- 1% of reading (nominal from calculation)
Pressure +/- 0.5% of scale
Other parameters Please consult factory
Sensitivity > 0.05ºC (0.09ºF)
Repeatability +/- 0.05ºC (+/- 0.09ºF)
Hysteresis None

Measurement Ranges

Chilled mirror sensors See table.
Temperature Sensors T-100, 80º to +100ºC (-112º to +212ºF)
Pressure Sensors

PT-30A, 0 to 30 psia (0 to 2 bar) -
PT-300A, 0 to 300 psia (0 to 21 bar) or
User-supplied analog signal.

Required Sample Flow 0.5 to 5.0 scfh (0.25 to 2.5 L/min)
Maximum Cooling Rate, Dew/Frost Point 1.5ºC (2.7ºF)/sec (typical, above 0ºC (32ºF))
Response Time
Temperature < 7 sec. (+25º to +70ºC (+77º to +158ºF) step change)
Pressure 1 sec to 90% of steady state (10% to 90% change)
Display Resolution Three displays, 6 digit, 0.1 or 0.01 selectable
Update Time 4 seconds max, RS-232C off. 8 seconds max, RS-232C on
 
Functional
Outputs
Analog Three 4-20 mADC, 500 ohm max. 0-5 VDC, 5mA max.
Analog Scaling (standard)
 
Dew/Frost Point Sensors 1111H, D-2: -40 to +60ºC (-40 to +140ºF)
SIM-12H: -25 to +100ºC (-13 to +212ºF)
1311DR, 1311XR: -80 to +20ºC (-112 to +68ºF)
Temperature -40Cº to +60Cº (-40º to +140Fº)
Pressure

PT-30A, 0 to 30 psia (0 to 2 bar)
PT-300A, 0 to 300 psia (0 to 21 bar)

Digital RS-232C
Alarm
Relay Three Form C, SPDT, 5A
Service Alarm TTL compatible. Service bit also available on RS-232C output.
PACER™ status TTL compatible. Status bit also available on RS-232C output.
Displays
Display 3 displays, 6 digit LED
Power 100, 115, 230 VAC (+15%, -20%) switch selectable, 50-60HZ, 75 watts max.
 
Operating Ranges
Sensor
Ambient Temperature -15º to +95ºC (+5º to +203ºF) depending on sensor
Pressure -14.7 to 300 psig (0 to 22 bar) depending on sensor
Electronics
Ambient Temperature 0º to +50ºC (+32º to +122ºF)
Relative Humidity 85% max
 
Physical
Weight 11kg. (23 lb.)
Shipping Weight 12 kg. (26lb.)
Environmental General purpose bench or rack mount.
 
Accessories
CC-M3 Carrying case
T-100 Temperature sensor
PT-30A Pressure sensor
PT-300A Pressure sensor