Ambient Temperature |
0º
to +50ºC (+32º to +122ºF) |
| Control
Temperature |
105ºC
(221º) |
| Material |
Sintered
stainless steel |
|
Pressure |
50
psig (4.5 bar) max. |
|
Voltage |
115/230
VAC, 50-60 Hz |
|
Power |
100
watts. |
| The
SIM-HFT filters the sample gas of particulate contaminants
prior to entering the SIM-12H sensor. The incoming
gas is first passed through a 90 micron pre-filter,
and then a 15 micron final filter. All parts in contact
with the gas sample are heated to a constant 105º
(221º) to prevent condensation. The filter elements
are easily removable for cleaning or replacement. |
| |
|
Ambient Temperature |
0º
to +50ºC (+32º to +122ºF) |
| Control
Temperature |
105ºC
(221º) |
| Material |
Glass,
aluminum |
|
Pressure |
50
psig (4.5 bar) max. |
|
Voltage |
115/230
VAC, 50-60 Hz |
|
Power |
75
watts. |
The
SIM-HFT controls the sample gas flow rate at the optimum
for the SIM-12H sensor. A front panel mounted metering
valve allows control over a range of 0 to 2 SCFH.
All parts in contact with the gas sample are heated
to a constant 105º C (221º F) to prevent
condensation. The SIM-HFM is mounted downstream of
the SIM-12H sensor.
|
| |
|
Ambient Temperature |
-15º
to +60ºC (+5º to +140Fº) |
| Control
Temperature |
125ºC
(257ºF) |
| Material |
Teflon |
|
Pressure |
50
psig (4.5 bar) max. |
|
Voltage |
115/230
VAC, 50-60 Hz |
|
Power |
250
watts. |
The
SIM-HSL is self-regulated at a temperature high enough
to ensure that no condensation will occur. It is available
in standard 15 ft. (4.6 m) lengths for 115 VAC, and
35 ft. (10.7m) lengths for 230 VAC. Several lengths
may be connected in series.
|
| |
| The
SIM-MPL accepts the SIM-12H, SIM-HFT, and SIM-HFM.
The mounting plate provides a convenient method of
wall mounting the entire heated sampling system. All
mounting, plumbing, and wiring is done at the factory,
providing a complete system ready for installation. |