| Generation
Method |
Divided
flow, Plane Surface Saturation |
|
Operating Range |
0º
to 40º |
| |
| Dewpoint
Generation Range |
-80º
to +15º at 25º ambient |
|
Adjustment |
Continuous |
| Repeatability |
+/-
0.25º |
| Accuracy
|
+/-
0.5º typical |
| Hysteresis |
None |
| Response
Time |
Less
than 3 min. |
| Stability |
+/-
0.25º |
| Storage |
-40º
to +50º |
Outputs
|
Constant
regulated flow; 0-5 SCFH (0-2.4 litres/min.) Max.
unregulated flow: 13 SCFH (6 litres/min.) |
| Input
Required |
Dry
gas less than 115 PPM (-40º dewpoint) 13 SCFH
(6 Litres/min.) 30-100 psig.
Even dryer gas is preferred as the dryness of the
inlet gas limits the lowest dewpoint that can be generated.
|
| |
| |
| Weight
|
20
pounds |
| Input/Output
Fittings |
1/4"
tubing compression fittings |
| |
| |
| Model
1500 or Hygro M2 with 1311DR or 1211H Sensor |
The
setup procedure for the DPG300 allows you to compute
the moisture content of your inlet gas utilizing the
generator and a precision chilled mirror hygrometer.
This dry gas calculation can be made even if the dry
gas moisture content is below the range of the dewpoint
sensor employed. |