| Generation
Method |
Divided
Flow |
| Operating
Range |
:
0 to 99% RH, 2º to 40ºC (35º to 100ºF) |
|
Accuracy |
0-10%
.......... +/- 4.0% RH
10-20% .......... +/- 2.0% RH
20-80% .......... +/- 1.0% RH
80-90% .......... +/- 2.0% RH
90-99.9% .......... +/-4.0% RH |
| Response
Time |
Less
than 15 minutes |
| Stability
|
0.2%
RH |
Storage |
| Full |
0.5º
to 40ºC (33º to 100ºF) |
| Drained |
-54º
to 50ºC (-65º to 120ºF) |
| |
| |
|
Power |
100/120/240
VAC, 50/60 Hz.*
Input power and frequency must be specified when ordering;
not field-changeable. |
|
Integral Power Supply |
24
VDC unregulated |
| |
| |
|
Weight |
12
lb. |
|
Dry Gas Inlet |
1/4"
tube fitting |
|
Chamber Dimensions |
81.3
x 147.3 x 94.0 mm (3.2"x5.8"x3.7"DP) |
| |
| |
| Power |
120
VAC 60 Hz |
| Flow
Rate |
1
to 2 SCFH |