C1/M - Overview
Portable Humidity Calibration Lab The
C1/M System is a NIST-traceable relative humidity calibration
chamber which is continuously monitored by a chilled mirror
hygrometer. The system utilizes the time proportioning divided
flow technique for generating precise % RH values. Set points
can be changed via the front panel from 10-95% RH. The system
acquires fast stability facilitating calibration cost-effective
calibration of humidity RH probes, transmitter and data
loggers. The modular design also provides portability and
enables the hygrometer to be used for validating humidity
and temperature conditions in environmental chambers, process
and storage areas.
The C1/M uses the divided flow method to
generate relative humidity values from 10-95% RH. The relative
humidity is controlled regulating a fraction of a constant
flow dry airstream, which is produced by an on-board desiccant
cartridge, through a saturator. The saturated air rejoins
the remaining dry air and mixes to the desired RH level.
Repeatable mixing of the dry air and saturated air is accomplished
by time-proportioning control. The relative humidity is
set via an increment/decrement switch and viewed on a bright
red LED display with 0.1% resolution. The C1 includes a
dry gas inlet that allows users with a dry gas source to
extend desiccant life indefinitely. The humidity of the
chamber is confirmed with an M series chilled mirror on
a second display. The hygrometer also provides analog and
digital outputs (RS-232).
The M series consists of the M2, M3, and
M4-RH hygrometers coupled to a chilled mirror sensor. These
instruments are fundamental primary dew point measurement
transfer standards. The M Series hygrometer samples air
from the C1 test chamber and measures dew point to ±0.2°C
or better uncertainty. A 100O platinum RTD measures the
C1 chamber temperature. From the primary dew point and temperature
readings, the M series hygrometer calculates and displays
percent relative humidity as well as provides and digital
signals for RH and temperature.
The system is equipped with an SST sampling
system which integrates the C1 and an M Series hygrometer.
It includes all necessary sampling components (pump, flow
restrictor orifice, chamber cover with tubing and fittings,
mounting plate for C1), as well as the necessary null modem
cable for connecting the C1 to the M Series hygrometer.
The M series hygrometer sends a constant stream of data
to the C1 via an RS232 interface. The C1 receives humidity
data from the hygrometer and continuously adjusts the time-proportioning
duty cycle to maintain the desired % RH set point. “RH
vs. Time Test Profiles” may be downloaded to the C1
from a PC by using the optional ProStep™ software.
The C1-M system is calibrated to NIST traceable
reference standards and is suitable for use as humidity
transfer standard. GE General Eastern’s manufacturing
and quality system is certified to ISO-9001 standards. For
more information on the theory and operation of GE General
Eastern’s Chilled Mirror Hygrometers and Humidity
Calibrators, contact us to obtain application assistance.
|