Sensing Element |
1/3
Class A DIN 43760 PRT, 100 ohm @ 0ºC (32ºF) |
Dew/Frost
Point Accuracy |
| Standard |
+/-
0.2º C(+/- 0.36ºF) |
| Optional |
+/-
0.15ºC (+/- 0.27ºF) |
| Sensitivity |
>
0.03ºC (0.05ºF) |
|
Repeatability |
+/-
0.05ºC (+/-0.09ºF) |
|
Hysteresis |
None |
|
Cooling Stages |
4 |
Depression
@ 25º,
1 atm,air |
|
| Air-cooled
|
95ºC
(171ºF) |
| Liquid-cooled
|
105ºC
(189ºF) |
| Air-cooled
|
95ºC
(171ºF) |
|
Typical Meas. Range @ 25ºC(77ºF),
1 atm, air |
|
Dew/Frost Point |
-65º
to +25ºC (-85º to +77ºF) |
| Relative
Humidity |
Equivalent
to 0.03% to 100% |
|
Liquid-cooled (15ºC (59ºF) cooling
water, at 25ºC (77ºF), 1 atm air) |
|
Dew/Frost Point |
-75º
to +15ºC (-103º to +59ºF) |
| Relative
Humidity |
Equivalent
to 0.007% to 100% |
|
Required Sample Flow |
0.5
to 5.0 scfh (0.25 to 2.5 L/min) |
| |
| |
| Power |
115/230
VAC, 50-60 Hz, 300 watts max. |
| |
| |
| Ambient
Temperature |
0º
to +35ºC(+32º to +95ºF) |
| Pressure |
0
to 300 psig (1 to 22 bar) |
| Auxiliary
Cooling |
Water
cooling (standard) |
| |
| |
|
Sensor Body Material |
Type
314 stainless steel |
|
Sensor Wetter Material |
316SS,
copper, mylar, silicone, rhodium or platinum mirror |
| Inlet/Outlet
Fittings |
1/4"
compression, 6 mm available |
| Weight |
16
kg. (34 lb.) |
| Shipping
Weight |
18
kg. (40 lb.) |
| |
| |
| HEX-1 |
Rhodium-plated
mirror |
| HEX-2 |
Solid
platinum mirror |
| TW-1 |
Mirror
torque wrench |
| 1123HK |
Sensor
cable (10 ft (3 m) standard), up to 300 ft (91 m)
available |